This IBV ELISA kit is intended Laboratory for Research use only and is not for use in diagnostic or therapeutic procedures.The Stop Solution changes the color from blue to yellow and the intensity of the color is measured at 450 nm using a spectrophotometer. In order to measure the concentration of IBV in the sample, this IBV ELISA Kit includes a set of calibration standards. The calibration standards are assayed at the same time as the samples and allow the operator to produce a standard curve of Optical Density versus IBV concentration. The concentration of IBV in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Samplecollection and storages
Serum - Use a serum separator tube and allow samples to clot for 30 minutes before centrifugation for 10 minutes at approximay 3000×g. Remove serum and assay immediay or aliquot and store samples at -20℃ or -80℃.Avoid repeated freeze-thaw cycles
Plasma- Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 30 minutes at 3000×g at 2-8℃ within 30 minutes of collection. Store samples at -20℃or -80℃. Avoid repeated freeze-thaw cycles.
Cell culture supernates and other biological fluids - Remove particulates by centrifugation and assay immediay or aliquot and store samples at -20℃or -80℃. Avoid repeated freeze-thaw cycles.
Note: The samples shoule be centrifugated dequay and no hemolysis or granule was allowed.
Materials required but not supplied
1. Standard microplate reader(450nm)
2. Precision pipettes and Disposable pipette tips.
3. 37 ℃ incubator
Precautions
1. Donot substitutereagentsfromone kit to another.Standard, conjugateandmicroplates are matchedfor optimal performance. Useonly thereagentssuppliedby manufacturer.
2. Donot removemicroplatefrom the storage baguntilneeded. Unusedstripsshouldbe stored at2-8°Cin their pouchwith the desiccantprovided.
3. Mix all reagents before using.
Remove allkit reagentsfrom refrigerator and allow them to reachroom temperature( 20-25°C)
Materials supplied
Name
96determinations
48determinations
Microelisa stripplate
96 strips
48 strips
Negative control
0.3ml
0.3ml
Positive control
0.3ml
0.3ml
Sample diluent
6.0ml
3.0ml
HRP-Conjugate reagent
10.0ml
5.0ml
20X Wash solution
25ml
15ml
Chromogen Solution A
6.0ml
3.0ml
Chromogen Solution B
6.0ml
3.0ml
Stop Solution
6.0ml
3.0ml
Closure plate membrane
2
2
User manual
1
1
Sealed bags
1
1
Reagent preparation
20×wash solution:Dilute with Distilled or deionized water 1:20.
Assay procedure
1. Prepare allreagentsbeforestartingassayprocedure. ItisrecommendedthatallStandardsand Samplesbe addedin duplicateto the MicroelisaStripplate.
2.Separay add Positive control and Negative control 50μl to the Positive and Negative well, Add testing sample 10μl Then add sample diluent 40μl to testing sample well; Blank well doesn’t add anyting.
3. Add100μlofHRP-conjugate reagent to each well,cover with an adhesive stripandincubatefor60 minutes at37°C.
4. Aspirate each well and wash, repeating the process four times for a total of five washes.Wash by filling each well with Wash Solution (400μl) using a squirt bottle, manifolddispenseror autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Solution by aspirating ordecanting. Invert the plate and blot it against clean paper towels.
5. Add chromogen solution A 50μl and chromogen solution B 50μl to each well. Gently mix and incubate for 15 minutes at 37°C. Protect from light.
6. Add 50μl Stop Solution to each well. The color in the wells should change from blue toyellow. If the color in the wells is green or the color change does not
appear uniform,gently tap the plate to ensure thorough mixing.